Teledyne API - T102/501 TRS, Addendum to T101 Operation Manual
T102 OPERATING INSTRUCTIONS
39
6.2.2.
T102 Test Channel Output
The following table supersedes the
Test Parameters Available for Analog Output A4
table in the
Test Channel Output
section in the T101 Manual - PN 07266
Table 6-3. Test Parameters Available for Analog Output A4
TEST CHANNEL
TEST PARAMETER RANGE
1
NONE
Test channel is turned off
PMT READING
0-5000 mV
UV READING
0-5000 mV
SAMPLE PRESSURE
0-40 in-Hg-A
SAMPLE FLOW
0-1000 cm³/min
RCELL TEMP
0-70° C
CHASSIS TEMP
0-70° C
IZS TEMP
0-70° C
PMT TEMP
0-50° C
CHASSIS TEMP
0-70° C
HVPS VOLTAGE
0-5000 V
1
This refers to the voltage range of the parameter and
not the output signal of the test channel.
Once a TEST function is selected, the instrument begins to report a signal on the
A36
output and adds
TEST=
to the list of test functions viewable on the display (just before
the TIME display).
6.3.
SETUP – COMM: Setting Up the T102’s Communication Ports
6.3.1.
T102 ID Code
The default ID code for all T102 analyzers is
102
.
To edit the instrument’s ID code, see
Analyzer ID
Section of the T101 Manual - PN
07266.
6.3.2.
T102 Ethernet Host Name
The default name for all TAPI Model T102 analyzers is
T102
.
To change the Ethernet Host Name see
Changing the Analyzer’s HOSTNAME
section in
the T101 Manual - PN 07266.
07267B DCN6485
Summary of Contents for 501 TRS
Page 2: ......
Page 4: ...ii This page intentionally left blank 07267B DCN6485 ...
Page 8: ...vi This page intentionally left blank 07267B DCN6485 ...
Page 10: ...viii This page intentionally left blank 07267B DCN6485 ...
Page 54: ...07267B DCN6485 ...
Page 92: ...Teledyne API Models T102 102E Software Documentation C5 05515B DCN5975 A 2 07267B DCN6485 ...
Page 98: ...This page intentionally left blank B 2 07267B DCN6485 ...
Page 102: ...This page intentionally left blank B 6 07267B DCN6485 ...
Page 105: ...APPENDIX D Wire List and Electronic Schematics ...
Page 106: ...This page intentionally left blank D 2 07267B DCN6485 ...
Page 110: ...D 6 07267B DCN6485 ...
Page 111: ......