Teledyne API - T102/501 TRS, Addendum to T101 Operation Manual
TROUBLESHOOTING and REPAIR
73
10.
Perform a lamp calibration procedure (see the Lamp Calibration Section
in T101 Manual - PN 07266) and a zero point and span point calibration
(see the Calibration Procedures section in the T101 Manual - PN 07266).
11.5.2.
Replacing the UV Filter/Lens
NOTE:
Be careful not to leave thumbprints on the interior of the sample chamber.
The various oils that make up fingerprints fluoresce brightly under UV light
and will significantly affect the accuracy of the analyzer’s SO
2
measurement)
1.
Turn off the instrument’s power and remove the power cord from
the instrument.
2.
Unplug the J4 connector from the motherboard to allow tool
access.
3.
Remove 4 screws from the shutter cover (see Figure 11-2) and
remove the cover.
4.
Remove 4 screws from the UV filter retainer (see Figure 11-2).
Figure 11-2. Disassembling the Shutter Assembly
5.
Carefully remove the UV filter.
07267B DCN6485
Summary of Contents for 501 TRS
Page 2: ......
Page 4: ...ii This page intentionally left blank 07267B DCN6485 ...
Page 8: ...vi This page intentionally left blank 07267B DCN6485 ...
Page 10: ...viii This page intentionally left blank 07267B DCN6485 ...
Page 54: ...07267B DCN6485 ...
Page 92: ...Teledyne API Models T102 102E Software Documentation C5 05515B DCN5975 A 2 07267B DCN6485 ...
Page 98: ...This page intentionally left blank B 2 07267B DCN6485 ...
Page 102: ...This page intentionally left blank B 6 07267B DCN6485 ...
Page 105: ...APPENDIX D Wire List and Electronic Schematics ...
Page 106: ...This page intentionally left blank D 2 07267B DCN6485 ...
Page 110: ...D 6 07267B DCN6485 ...
Page 111: ......