"V VARIABLE=VALUE WARNLO WARNHI [DATALO-DATAHI] <CRLF>"
which should reflect the new value. The values in square brackets are not required for all
variables. If needed, the values are included on the command line, separated by spaces.
For example, to change the instrument ID, the host computer would issue a command like
this:
"V MACHINE_ID=1234<CRLF>"
and the CPU should respond with:
"V DDD:HH:MM IIII MACHINE_ID = 1234 (0-9999)<CRLF>"
Table 7.9 lists the variable names which are variable through the RS-232 interface and
their corresponding button sequences.
RS-232 VARIABLE NAMES
Var. Name
Button Sequence
Legal Values
CO_SPAN
SETUP-IZSC-SPAN
0-1000
FILT_ADAPT
SETUP-MISC-FILT-ADAPT
OFF, ON
REPORT_FREQ
SETUP-DAS-FREQ
1-60
MACHINE_ID
SETUP-COMM-ID
0000-9999
BAUD_RATE
SETUP-COMM-BAUD
300,1200,2300
CURR_TIME
SETUP-MISC-CLK-TIME
00:00-23:59
CURR_DATE
SETUP-MISC-CLK-DATE
01/01/00-12/31/99
CLOCK_ADJ
SETUP-MISC-CLK-ADJ
-60 - +60
FAULT_TIME SETUP-MISC-MORE-FLT
0-300
TABLE 7.9