P/N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual - Page 44
Triggering Events
Name
Description
ATIMER
Automatic timer expired
EXITZR
Exit zero calibration
EXITSP
Exit span calibration
EXITMP
Exit multi-point calibration
SLPCHG Slope
changed
REPORT PERIOD
This property specifies the period between reports (DAS entries). Enter the number of
days and press
ENTR
. Then enter the number of hours (the leftmost digits) and minutes
and press
ENTR
.
NUMBER OF RECORDS
This property specifies how many records you want to store in this data channel. Each
record consists of a time stamp and the sampled measurements from each parameter. The
memory of the M300 has capacity for a total of about 8,000 records (for
all
data
channels). When the specified number of records for a data channel has been reached,
the channel will “wrap around”, with the oldest records being overwritten with the newest
records.
Enter the number of records desired and press
ENTR
. Only when you press
ENTR
is
the data file actually deleted. If you press
EXIT
the property will not be modified and
data file will not be deleted.
RS-232 REPORT
This property controls printing of reports on the RS-232 channel. If set to ON then every
time a new report is stored it’s also printed on the RS-232 channel. If this property is set
to OFF, the reports are still stored in the data file, but not printed on the RS-232 channel.
PARAMETERS
This property shows the
number
of parameters being sampled. Each data channel can
sample up to 10 parameters, each with a separate sampling mode.
This parameter also specifies the
mode
of data collection. When sampling a reading such
as a concentration at periodic intervals, it is desirable to use the
AVG mode
to generate an
average of the readings during the interval. But when sampling the concentration when
NOTE
Changing this property requires that any existing data records
be deleted because memory must be re-initialized.