P/N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual - Page 27
ILLUSTRATION OF NORMAL DISPLAY
FIGURE 2.2
If
TEST
is pushed, the upper center display cycles through the menu of test parameters,
e.g. Sample flow (see Table 2.3). If
CALZ
is pushed, the sequence of operations for
setting the Analyzer zero is initiated(see Section 3.1).
If
CALS
is pushed, the sequence of operations for setting
SPAN
is initiated (see Section
3.1).
CAL
is used to initiate span setting using sample gas, such as during a formal
calibration.
Pushing
MSG
will cause a message to appear on the upper center display.
Pushing
CLR
will erase a message, provided the condition causing the message has
ceased.
Pushing
SETUP
changes the function of the push buttons and is used for setting basic
parameters as described in Section 4.0.
2.2.3 Status LED’s
The three status
LED’s
to the right of the display indicate the general status of the Model
300 Analyzer. The green
SAMPLE LED
indicates the sampling status. The yellow
CAL LED
indicates the calibration status. The red
FAULT LED
indicates the fault
status. Table 2.5 summarizes the meanings of the status
LED’s
.