
P/N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual - Page 32
minute (00 - 59). Enter the time of day for calibration check and then press
ENTR
to
accept the new time or
EXIT
to leave the time unchanged.
Delta Days:
The number of delta days is the number of days between each auto-
sequence. Enter desired number of delay days(0-365) and press
ENTR
.
Delta Time:
The delta time allows the automatic Z/S check time-of-day to be delayed in
the format of HH:MM, where HH is the hour from 00 to 23 and MM is the minutes(00-
59). The delta days and delta time are added together to determine the total delay
between sequences. The delta time parameter allows you to advance or retard the starting
time by a fixed amount each time the sequence is run. For example: Setting the delta
days to 1 day and the delta time to 15 minutes will delay the starting time for the
sequence by 15 minutes each day. If you want to have the sequence run at the same time
every day, simply set the delta time to zero.
Duration:
The duration of each step of the sequence. Enter the duration in minutes(1-60)
and press
ENTR
.
Calibrate:
When set to
ON
, the sequence will adjust the internal formulas (slope and
offset) in the Analyzer to the value set in the span variable. If this feature is enabled
along with the automatic zero/span check, the Analyzer will re-adjust its formulas to
match the predetermined zero and span settings once each day.
NOTE
THE PROGRAMMED START TIME MUST BE A
MINIMUM OF 5 MINUTES LATER THAN THE REAL
TIME CLOCK (See Section 4.3 and 4.4 for setting real time
clock).
NOTE
Avoid setting two or more sequences at the same time of the
day. Any new sequence which is initiated whether from a
timer, the RS-232, or the contact closure inputs will override
any sequence which is in progress.