exiting calibration it is desirable to take a single instantaneous reading using the
INST
mode
. The table below summarizes the available sampling modes.
Sampling Modes
Mode
Description
INST Records
instantaneous
reading
AVG
Records average reading during reporting interval
MIN
Records minimum reading during reporting interval
MAX
Records maximum reading during reporting interval
Each parameter in this list can refer to a different (or the same) point, and can use a
different (or the same) sampling mode. For instance, you could create three parameters
which all measure the CO concentration for range 1 (the CONC1 parameter), but which
use the MIN, AVG, and MAX sampling modes. This permits you to record not only the
average concentration during the reporting interval, but the maximum excursions as well.
Like data channels, parameters also have properties which are listed in the table below.
Parameter Properties
Property
Description
Initial
Setting
Setting Range
PARAMETER
The parameter to sample
CONC1
Any of the parameters
listed in the table of
parameters shown
above
SAMPLE
MODE
The sampling mode to use
when reading this
parameter
AVG
Any one of the INST,
AVG, MIN, and MAX.
PRECISION
Number of decimal digits
1
0 - 3
The table below lists the data parameters initially defined. The entries in the Name and
Units columns are visible when editing and printing the data channel properties.