P/N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual - Page 34
4. STARTING TIME: 23:30
5. DELTA DAYS: 7
6. DELTA TIME: 00:00
7. DURATION: 15
8. CALIBRATE: OFF
Example 4
: to perform zero check once per day at 10:30 pm and a span check once per
week starting at 11:30 pm, 12/20/93.
1. Select any one of SEQx and program as example 1.
2. Select any other SEQx and program as example 3. Avoid setting
two or more sequences at the same time of the day.
3.5 Summary of front panel check and calibration
controls
The calibration controls are summarized below in terms of the button sequences used to
access them.
CALIBRATION CONTROLS
Button Sequence
Function
Default
Limits
CALZ
Begin zero check
-
-
CALZ-ZERO-ENTR
Adjust CO conc. to zero
-
-
CALZ-EXIT
Exit zero check
-
-
CALS Begin
span
check
-
-
CALS-CONC-ENTR
Expected CO span value
40 PPM
1-1000 PPM
CALS-SPAN-ENTR
Adj. CO conc. to span val.
-
-
CALS-EXIT
Exit span check
-
-
CAL
Begin M-P cal.
-
-
CAL-ZERO-ENTR
Adj. CO conc. zero value
-
-
CAL-CONC-ENTR
Expected CO span value
40 PPM
1-1000 PPM
CAL-SPAN-ENTR
Adj. CO conc. to span val.
-
-
CAL-EXIT
Exit M-P cal.
-
-
SETUP-ACAL-SEQ1
Setup auto-cal SEQ1
Disabled
SETUP-ACAL-SEQ2
Setup auto-cal SEQ2
Disabled
SETUP-ACAL-SEQ3
Setup auto-cal SEQ3
Disabled
TABLE 3.1