Behavior
When you select A Only with Delay Mode on, the instrument first looks for the trigger conditions
specified by the A Event. The instrument then delays the acquisition for the time specified by the
Horiz Delay control before acquiring the data.
Trigger After Time (Horizontal Delay Off)
To Use
Set up the A Event and B Event from their individual tabs. Select the A–
>
B Seq tab and click Trig
After Time. Select the trigger delay time with the Trig Delay control. Adjust the Horiz Position
and B Trig Level controls as necessary.
Behavior
When you select Trig After Time, the instrument looks for the trigger conditions specified by the
A event. After the A event arms the trigger system, the instrument triggers on the next B event that
occurs after the time specified by the Trig Delay control.