117
Select User Preferences from the Utilities menu to display the Prompt Before Action window.
Click Autoset to toggle between OFF and ON:
OFF to immediately perform an Autoset when the AUTOSET button is pushed.
ON to display a prompt when the AUTOSET button is pushed.
Click Close to save your selection.
The following table lists the Autoset defaults.
Control
Changed by Autoset to
Selected Channel
Numerically lowest displayed channel
Acquisition Mode
Sample
RUN/STOP RUN
Deskew, Channel/Probe
Unchanged
Display Style
Vectors
Display Intensity—Overall
If less than 50%, set to 100%
Display Format
YT
Horizontal Position
Centered within the graticule window
Graticule Setting
Unchanged
AutoBright On
Horizontal Scale
As determined by the signal frequency,
or 100
µ
s/div
Record Length and
Sample Rate
When a new time-per-division is
calculated, the sample rate changes to
maintain the current record length. If the
sample rate cannot be changed because
of hardware limitations, the sample rate
is set to maximum while the record
length is reduced to maintain the time-
per-division.
Trigger Position
Delay Off
Trigger A >B Sequence
A Only
Main Trigger Type
Edge
Trigger Source
Numerically lowest displayed channel
(the selected channel)
Trigger Level
50% of selected channel’s amplitude
Trigger Slope
Positive