327
If you are using H Bars or V Bars there is only one source
for both Cursor 1 and 2. If you are using Waveform or
Screen cursors, select a source for both Cursor 1 and 2.
If a waveform is not available, it appears shaded in the drop-down
list.
Change the cursor
position
5.
To change the position of a cursor, use the multipurpose
knobs, enter a numeric value in the Curs# Pos controls,
or click and drag the cursors to a new position.
Set cursor tracking
6.
To change the cursor tracking mode, click Setup on the
Cursor Type Setup control window. The Cursor Setup
control window opens. (All adjustments that you can make
in the Cursor Type Setup control window can also be made
in this window.)
7.
Select a cursor Track Mode:
Indep makes each cursor moveable without regard to
the position of the other cursor.
Tracking makes both cursors move in unison and
maintain a fixed horizontal or vertical distance between
each other.
8.
To turn the display of the cursors on or off, push the
front-panel Cursor button.
Set cursor style
9.
For waveform or screen cursors set the cursor style by
selecting one of the following from the list:
Lines display as the intersection of horizontal and
vertical lines. Lines are the default style for Screen
cursors.
Line & X display the horizontal position as a vertical
line; the Y-axis is displayed as an X.
X displays the horizontal and vertical position as a
single X.
Starting and Resetting Histogram Counting
Use this procedure to quickly take a measurement based on the default settings for histograms.
Prerequisites
1.
Display the waveform(s) to be measured on the screen.
Open histogram setup
window
2.
From the toolbar, click the Meas button to open the
Measurement Setup control window. Click Histogram to
open the Histogram control window.