Adjustments
Adjust the 30 A Range
4.
Set the probe range to 30 A.
5.
Set the calibrator output to 3.0 A.
6.
Adjust the 30 A DC gain control in the probe
compensation box to display 1.50 V, ±0.045 V on the
DMM.
7.
Disconnect the probe from the current source.
66
TCP0030A Current Probe Instruction Manual