Performance Veri
fi
cation
Equipment Required
The equipment required for the performance veri
fi
cation procedure is listed below. (See Table 6.)
Table 6: Test equipment
Description and quantity
Performance requirement
Recommended example
1
Oscilloscope
TekVPI interface, 500 MHz or greater bandwidth
Tektronix DPO4000
High amplitude pulse generator
Risetime <500 ps, pulse width >100 ns, amplitude
>10 Vpp into 50
Ω
Picosecond Labs 2600C
Calibrator
DCV: 0.2% accuracy, 0 to ±1.5 V, square wave output
ACA: 0.25% accuracy, 0 to ±6 A, square wave output
Fluke 9100
DMM
≥
2 volt range,
≥
1% accuracy
Keithley 2700
TekVPI Calibration/Veri
fi
cation
adapter
TekVPI interface
067-1701-xx (See page 42.)
DC current loop
5 turns 18 AWG coated wire on 3 inch form
See instructions that follow
HF current loop
50
Ω
±0.5%, BNC male
067-2396-xx
Adapter
BNC-to-dual banana
103-0090-xx
Adapter
SMA male-to-BNC female
015-1018-xx
BNC cable
50
Ω
, 0.76 m (30 in) length
012-0117-xx
Test leads (2)
Banana plug to clip leads
Keithley 1681
1
Nine-digit part numbers (xxx-xxxx-xx) are Tektronix part numbers.
54
TCP0030A Current Probe Instruction Manual
Summary of Contents for TCP0030A
Page 1: ...TCP0030A 120 MHz 30 Amp AC DC Current Probe ZZZ Instruction Manual xx www tek com 071 3006 01...
Page 4: ......
Page 8: ...Table of Contents iv TCP0030A Current Probe Instruction Manual...
Page 20: ...Preface xvi TCP0030A Current Probe Instruction Manual...
Page 71: ...Specifications TCP0030A Current Probe Instruction Manual 51...