Performance Veri
fi
cation
30 A Range Test
17.
Set the current source output to +3.00 A.
18.
Set the probe to the 30 A range.
19.
Record the exact measurement of the
DMM as M1.
20.
Set the current source output to -3.00 A.
21.
Record the exact measurement of the
DMM as M2.
22.
Compute the % Error using the measured
amplitude values and the formula shown.
23.
Record the computed %Error values in the
Test Record.
24.
Disable the calibrator output.
Test current = ±15.00 A (±3.00 A source output x 5 coil turns)
Expected output voltage (Ve) = 1.50 V