Adjustments
DC Gain Accuracy
This procedure describes the DC gain accuracy adjustments for the probe. You must
fi
rst adjust the 5 A range, and then adjust 30 A
range, in that order.
Adjust the 5 A Range
1.
Complete the
DC Gain Accuracy
test in the
Performance Veri
fi
cation
procedure. (See page 57,
.) If the measurements fall outside
the test limits, proceed to the next step.
2.
Set the calibrator output to 0.5 A.
3.
Adjust the 5 A DC gain control in the probe head to
display 2.50 V, ±0.075 V on the DMM.
TCP0030A Current Probe Instruction Manual
65