Performance Veri
fi
cation
6.
Set the probe range to the 5 A setting.
7.
Degauss the probe.
8.
Clamp the current probe around the HF
current loop. Verify that the arrow-shaped
indicator on the probe points away from the
pulse generator.
9.
Record the rise time measurement in the
Test Record.
NOTE.
If the measured rise time is out of
tolerance, you may be able to readjust it. (See
page 64, Adjustments.)
10.
Calculate the probe bandwidth using the
measured rise time in the following formula:
11.
Record the calculated bandwidth value in
the test record.
12.
Set the probe range to the 30 A setting and
set the vertical sensitivity to 200 mA/div.
13.
Remove the probe from the current loop
and then repeat steps 7 through 11.
End of procedures.
62
TCP0030A Current Probe Instruction Manual
Summary of Contents for TCP0030A
Page 1: ...TCP0030A 120 MHz 30 Amp AC DC Current Probe ZZZ Instruction Manual xx www tek com 071 3006 01...
Page 4: ......
Page 8: ...Table of Contents iv TCP0030A Current Probe Instruction Manual...
Page 20: ...Preface xvi TCP0030A Current Probe Instruction Manual...
Page 71: ...Specifications TCP0030A Current Probe Instruction Manual 51...