Performance Veri
fi
cation
Equipment Setup
Use the following procedure to set up and warm up the equipment to test the probe.
1.
Turn on the oscilloscope.
2.
Connect the probe to any channel (1–4) of
the oscilloscope.
3.
Close the probe slider and then press the
Degauss/AutoZero button.
4.
Set the oscilloscope coupling to DC.
5.
Power on the current source and the pulse
generator.
6.
Allow 20 minutes for the equipment to
warm up.
7.
Photocopy the test record and use it to
record the test results. (See page 63.)
56
TCP0030A Current Probe Instruction Manual
Summary of Contents for TCP0030A
Page 1: ...TCP0030A 120 MHz 30 Amp AC DC Current Probe ZZZ Instruction Manual xx www tek com 071 3006 01...
Page 4: ......
Page 8: ...Table of Contents iv TCP0030A Current Probe Instruction Manual...
Page 20: ...Preface xvi TCP0030A Current Probe Instruction Manual...
Page 71: ...Specifications TCP0030A Current Probe Instruction Manual 51...