Adjustments
High Frequency Response
Use this procedure to bring the rise time of the probe to within the warranted speci
fi
cations. This adjustment also affects the
aberrations on the probe output waveform, so for each current range, you may need to optimize the
fi
nal adjustment between the
two characteristics.
1.
Complete the
Rise Time and Bandwidth
test in the
Performance Veri
fi
cation
procedure. If the measurements fall
outside the test limits, proceed to the next
step.
2.
Set the Range to 5 A.
3.
Adjust the H.F. control to optimize the rise
time and aberrations.
4.
Set the Range to 30 A and repeat step 3.
5.
Check that the probe response is optimized
for both ranges.
TCP0030A Current Probe Instruction Manual
67