Setting channel input parameters
Setting channel input parameters
Use the vertical
Menu
buttons to select waveforms to display or open menus and submenus with which to set input
parameters for each channel.
Each channel’s settings are independent of every other channel.
Setting input signal coupling
Input signal coupling sets how the input signal is passed to the oscilloscope sampling circuit.
1.
Push the
Vertical Menu
button for the
channel input that you want to set.
2.
Push the
Coupling
side-menu button.
3.
Use the
Multipurpose
knob to select and click the coupling type:
DC
coupling passes both AC and DC signal components.
AC
coupling blocks the DC component and passes only the AC signal.
Ground
internally disconnects the input signal and connects that channel to a ground reference. This is a fast way to
determine where the ground level is for signals that have a DC component or offset.
Inverting the input signal
Use this procedure to invert (
fl
ip vertically) the signal. A typical reason to invert a signal is to use the inverted signal to
create a math waveform.
1.
Push the
Vertical Menu
button for the
channel input that you want to set.
2.
Push the
Invert
side-menu button to toggle between
Off
for normal operation and
On
to invert the polarity of the signal
in the preampli
fi
er.