The default oscilloscope settings (Default Setup)
The default oscilloscope settings (Default Setup)
The following table lists the oscilloscope settings that are applied when you push the
Default Setup
button.
NOTE.
When you push the
Default Setup
button, the oscilloscope displays the channel 1 waveform and removes all other
waveforms.
Function
Setting and value
Acquire
Mode:
Sample
Record length
: 2000 points
Cursors
Off
Source
: Ch1
Display
Persist
:
Auto
Mode: YT
Backlight
: 80%
Graticule
:
On
FFT
Source
wfm:
On
Vertical
Units: dBv RMS
Window
:
Hanning
Horizontal
Scale
(time per major horizontal division): 4.00
μ
s/div
Delay
:
On
Trigger
position: 0.00 s
Math
waveform
Source
1: Ch1
Source
2: Ch2
Operation
: + (add)
Measurements
Source
: Ch1
Selected measurements: None
Trigger
Type
:
Edge
Source
:
Ch1
Coupling
:
DC
Slope
:
Rising
Level
: 0.00 V
Mode:
Auto
USB Device Port
Connect to Computer
Vertical
(all channels)
Coupling
:
DC
Invert
:
Off
Bandwidth
:
Full
Voltage probe attenuation: 10X
Baseline position: 0.00 V
Scale
(volts per major vertical division): 1.00 V