Recommended calibration parameters
Note: Parameter steps for each range may be performed in any order, but all three parameter
steps for each range must be completed. For optimum calibration, use parameters within
recommended limits.
The maximum calibration command parameter ranges are: 75% to 150%
of full scale for positive and negative full scale calibration points; ± zero cal-
ibration steps have ±50% of full scale for valid entry ranges. However, for
optimum calibration, use calibration points within the ranges listed in Table
2-6 and Table 2-7. Note that each sense range requires three parameters:
zero, negative full scale, and positive full scale. Similarly, each source range
requires four parameters: two zero parameters, a positive full-scale
parameter, and a negative full-scale parameter.
Table 2-6
Recommended :CALibration:PROTected:SENSe parameter ranges
Sense
range
First parameter
(zero)
Second parameter
(negative full scale)
Third parameter
(positive full scale)
0.2V
2V
20V
1000V
1µA
10µA
100µA
1mA
20mA
100mA
1A
-0.002 to +0.002
-0.02 to +0.02
-0.2 to +0.2
-2 to +2
-1E-8 to +1E-8
-1E-7 to +1E-7
-1E-6 to +1E-6
-1E-5 to +1E-5
-1E-4 to +1E-4
-1E-3 to +1E-3
-1E-2 to +1E-2
-0.18 to -0.22
-1.8 to -2.2
-18 to -22
-800 to -1100
-0.9E-6 to -1.1E-6
-9E-6 to -11E-6
-90E-6 to -110E-6
-0.9E-3 to -1.1E-3
-18E-3 to -22E-3
-90E-3 to -110E-3
-0.9 to -1.1
+0.18 to +0.22
+1.8 to +2.2
+18 to +22
+800 to +1100
+0.9E-6 to +1.1E-6
+9E-6 to +11E-6
+90E-6 to +110E-6
+0.9E-3 to +1.1E-3
+18E-3 to +22E-3
+90E-3 to +110E-3
+0.9 to +1.1
Calibration
2-17
2410-902-01.book Page 17 Monday, November 7, 2005 2:49 PM
Summary of Contents for Keithley SourceMeter 2410
Page 11: ...1 Performance Verification 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 29: ...2 Calibration 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 53: ...3 Routine Maintenance 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 56: ...4 Troubleshooting 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 74: ...5 Disassembly 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 84: ...6 Replaceable Parts 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 100: ...A Specifications 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 102: ...B Command Reference 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 116: ...C Calibration Programs 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...