:SENSe
(:CALibration:PROTected:SENSe)
Purpose
To calibrate the sense function.
Format
:cal:prot:sens <nrf>
Parameters
See Table B-2 for recommended parameter ranges for optimum calibration.
Description
The :CAL:PROT:SENS command calibrates the Model 2410 sense function.
To use this command, you must (1) program the source to the correct value,
(2) select the range being calibrated, and (3) send the :CAL:PROT:SENS
command for each parameter listed in Table B-2.
Note
When the Model 2410 receives this command, it will attempt to place the
parameter into one of the three parameter ranges summarized in Table B-2,
depending on the active :SENS:FUNC and range. If the parameter does not
fit into any of the three allowed ranges, error number -222, "Parameter data
out of range," will be generated. Once the unit has successfully selected the
appropriate parameter range, it will then check to see if autorange for the
active :SENS:FUNC is enabled. If so, error number -221, "Settings conflict,"
will be generated. If no error occurs, the active sense function/range point
will be calibrated using the corresponding parameter.
Note that parameters for a given sense function and range can be sent in any
order. However, once one :CAL:PROT:SENS command executes, the other
two must also be executed or error number -200, "Execution error," will
occur when the :CAL:PROT:SAVE command is processed at the end of the
calibration procedure.
Example
:CAL:PROT:SENS 2
Calibrate 2V sense range.
B-6
Command Reference
2410-902-01.book Page 6 Monday, November 7, 2005 2:49 PM
Summary of Contents for Keithley SourceMeter 2410
Page 11: ...1 Performance Verification 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 29: ...2 Calibration 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 53: ...3 Routine Maintenance 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 56: ...4 Troubleshooting 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 74: ...5 Disassembly 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 84: ...6 Replaceable Parts 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 100: ...A Specifications 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 102: ...B Command Reference 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 116: ...C Calibration Programs 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...