![Tektronix Keithley SourceMeter 2410 Service Manual Download Page 113](http://html1.mh-extra.com/html/tektronix/keithley-sourcemeter-2410/keithley-sourcemeter-2410_service-manual_1078031113.webp)
Status byte EAV (Error Available) bit
Whenever an error is available in the error queue, the EAV (Error Available) bit (bit 2) of the
status byte will be set. Use the *STB? query to obtain the status byte, and then test bit 2 to see
if it is set. If the EAV bit is set, an error has occurred, and you can use the appropriate error query
to read the error and at the same time clear the EAV bit in the status byte.
Generating an SRQ on error
To program the instrument to generate an IEEE-488 bus SRQ (Service Request) when an
error occurs, send the *SRE 4 command. This command will enable SRQ when the EAV bit is
set. You can then read the status byte and error queue as outlined above to check for errors and
to determine the exact nature of the error.
Table B-4
Calibration errors
Error
number
Error message
+500
+501
+502
+503
+504
+505
+506
+507
+508
"Date of calibration not set"
"Next date of calibration not set"
"Calibration data invalid"
"DAC calibration overflow"
"DAC calibration underflow"
"Source offset data invalid"
"Source gain data invalid"
"Measurement offset data invalid"
"Measurement gain data invalid"
B-12
Command Reference
2410-902-01.book Page 12 Monday, November 7, 2005 2:49 PM
Summary of Contents for Keithley SourceMeter 2410
Page 11: ...1 Performance Verification 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 29: ...2 Calibration 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 53: ...3 Routine Maintenance 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 56: ...4 Troubleshooting 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 74: ...5 Disassembly 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 84: ...6 Replaceable Parts 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 100: ...A Specifications 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 102: ...B Command Reference 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 116: ...C Calibration Programs 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...