Step 4: Enter calibration dates and save calibration
NOTE
For temporary calibration without saving new calibration constants, proceed to Step
5: Lock out calibration.
1. From normal display, press MENU.
2. Select CAL, and then press ENTER. The Model 2410 will display the following:
CALIBRATION
UNLOCK EXECUTE VIEW-DATES
SAVE LOCK CHANGE-PASSWORD
3.
Select SAVE, and then press ENTER. The instrument will display the following
message:
SAVE CAL
Press ENTER to continue; EXIT to abort calibration sequence.
4.
Press ENTER. The unit will prompt you for the calibration date:
CAL DATE:
4
/15/
96
Use
,
,
▲
,
▼
,
ENTER, or EXIT.
5.
Change the displayed date to today's date, and then press the ENTER key. Press
ENTER again to confirm the date.
6.
The unit will then prompt for the calibration due date:
NEXT CAL:
4
/15/
97
Use
,
,
▲
,
▼
, ENTER, or EXIT.
7.
Set the calibration due date to the desired value, and then press ENTER. Press ENTER
again to confirm the date.
8.
Once the calibration dates are entered, calibration is complete and the following message
will be displayed:
CALIBRATION COMPLETE
Press ENTER to confirm; EXIT to abort
9.
Press ENTER to save the calibration data (or press EXIT to abort without saving
calibration data). The following message will be displayed:
CALIBRATION SUCCESS
Press ENTER or EXIT to continue.
10.
Press ENTER or EXIT to complete process.
Step 5: Lock out calibration
1. From normal display, press MENU.
2. Select CAL, and then press ENTER. The Model 2410 will display the following:
CALIBRATION
UNLOCK EXECUTE VIEW-DATES
SAVE LOCK CHANGE-PASSWORD
3.
Select LOCK, and then press ENTER. The instrument will display the following
message:
CALIBRATION LOCKED.
Press ENTER or EXIT to continue.
4.
Press ENTER or EXIT to return to normal display.
Calibration
2-15
2410-902-01.book Page 15 Monday, November 7, 2005 2:49 PM
Summary of Contents for Keithley SourceMeter 2410
Page 11: ...1 Performance Verification 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 29: ...2 Calibration 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 53: ...3 Routine Maintenance 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 56: ...4 Troubleshooting 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 74: ...5 Disassembly 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 84: ...6 Replaceable Parts 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 100: ...A Specifications 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 102: ...B Command Reference 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 116: ...C Calibration Programs 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...