Table 2-3
Front panel voltage calibration
Source
range*
Source
voltage
Multimeter voltage
reading**
0.2V
2V
20V
1000V
+200.00mV
+000.00mV
-200.00mV
-000.00mV
+2.0000V
+0.0000V
-2.0000V
-0.0000V
+20.000V
+00.000V
-20.000V
-00.000V
+975.00V
+000.00V
-975.00V
-000.00V
___________ mV
___________ mV
___________ mV
___________ mV
_____________V
_____________V
_____________V
_____________V
_____________V
_____________V
_____________V
_____________V
_____________V
_____________V
_____________V
_____________V
*
*Use EDIT and RANGE keys to select source range.
**Multimeter reading used in corresponding calibration
step. See procedure.
Model 2410
Digital Multimeter
Input HI
Input LO
2410
SourceMeter
250V
PEAK
5V
PEAK
HI
LO
OUTPUT
1100V
PEAK
1100V
PEAK
EDIT
TOGGLE
POWER
RANGE
INPUT/
OUTPUT
SENSE
Ω
4 WIRE
DISPLAY
ON/OFF
TERMINALS
FRONT/
REAR
AUTO
RANGE
!
EXIT
ENTER
CONFIG MENU
SWEEP
TRIG
REL
LOCAL
FILTER
LIMIT
DIGITS SPEED
V
Ω
MEAS
I
FCTN
V
I
SOURCE
2
3
0
1
6
7
8
9
4
+/-
5
STORE RECALL
EDIT
4-WIRE
SENSE
2410
1100V SourceMeter
®
!
Figure 2-1
Voltage
calibration
test connections
Calibration
2-11
2410-902-01.book Page 11 Monday, November 7, 2005 2:49 PM
Summary of Contents for Keithley SourceMeter 2410
Page 11: ...1 Performance Verification 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 29: ...2 Calibration 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 53: ...3 Routine Maintenance 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 56: ...4 Troubleshooting 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 74: ...5 Disassembly 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 84: ...6 Replaceable Parts 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 100: ...A Specifications 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 102: ...B Command Reference 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 116: ...C Calibration Programs 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...