![Tektronix Keithley SourceMeter 2410 Service Manual Download Page 71](http://html1.mh-extra.com/html/tektronix/keithley-sourcemeter-2410/keithley-sourcemeter-2410_service-manual_1078031071.webp)
Analog circuitry checks
Table 4-4 summarizes analog circuitry checks.
*
* Measured with respect to FCOM (TP500).
** Measured with respect to OCOM (TP501).
Table 4-4
Analog circuitry checks
Step
Item/
component Required condition
Remarks
1*
2*
3*
4*
5*
6**
7**
8**
9*
TP200
TP201
TP202
TP203
TP213
TP218
TP219
TP214
TP232
>1100V voltage protection
10V
10V (SVMI)
10V
10V
OUTPUT COM
OUTPUT COM
SVMI, OUTPUT ON, 20V
Bench defaults
-13V ±1V
-5V ±.5V
-10V ±1V
-10.5V ±1V
0V ±.1V
7V ±.7V
7V ±.7V
20V ±.5V
6.4V ±.6V
4-16
Troubleshooting
2410-902-01.book Page 16 Monday, November 7, 2005 2:49 PM
Summary of Contents for Keithley SourceMeter 2410
Page 11: ...1 Performance Verification 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 29: ...2 Calibration 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 53: ...3 Routine Maintenance 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 56: ...4 Troubleshooting 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 74: ...5 Disassembly 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 84: ...6 Replaceable Parts 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 100: ...A Specifications 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 102: ...B Command Reference 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 116: ...C Calibration Programs 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...