Remote calibration
Use the following procedure to perform remote calibration by sending SCPI commands over
the IEEE-488 bus or RS-232 port. The remote commands and appropriate parameters are
separately summarized for each step.
Remote calibration commands
Table 2-5 summarizes remote calibration commands. For a more complete description of
these commands, refer to Appendix B.
*Calibration data will not be saved if:
1. Calibration was not unlocked with :CODE command.
2. Invalid data exists. (For example, cal step failed or was aborted.)
3. Incomplete number of cal steps were performed. (For example, omitting a negative full-scale step.)
Ranges that successfully calibrated will be saved if calibration is unlocked. Ranges that failed will not be
saved.
Table 2-5
Remote calibration command summary
Command
Description
:CALibration
:PROTected
:CODE '<password>'
:COUNT?
:SAVE
:LOCK
:LOCK?
:DATE <y>,<m>,<d>
:DATE?
:NDUE <y>,<m>,<d>
:NDUE?
:SENSe <nrf>
:DATA?
:SOURce <nrf>
:DATA?
Calibration subsystem.
Cal commands protected by password.
Unlock cal; changes password if cal is already unlocked.
(Default password: KI002410.)
Query number of times 2410 has been calibrated.
Save calibration data to EEPROM.*
Lock calibration, inhibit SAVE command operation.
Request cal lock status.
Program calibration year, month, day.
Query calibration year, month, day.
Program calibration due year, month, day.
Query calibration due year, month, day.
Calibrate active measure range. (See Table 2-6 parameters.)
Query measurement cal constants for active range.
Calibrate active source range. (See Table 2-7 parameters.)
Query source cal constants for active range.
2-16
Calibration
2410-902-01.book Page 16 Monday, November 7, 2005 2:49 PM
Summary of Contents for Keithley SourceMeter 2410
Page 11: ...1 Performance Verification 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 29: ...2 Calibration 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 53: ...3 Routine Maintenance 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 56: ...4 Troubleshooting 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 74: ...5 Disassembly 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 84: ...6 Replaceable Parts 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 100: ...A Specifications 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 102: ...B Command Reference 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 116: ...C Calibration Programs 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...