*1. Perform complete procedure for each range, where <Range> and <Source_value> = 1E-6, 10E-6, 100E-6,
1E-3, 20E-3, 100E-3, or 1.
2. <DMM_Reading> parameter is the multimeter reading from the previous step.
3. Use :SYST:ERR? query to check for errors.
Table 2-10
Current calibration initialization commands
Command
Description
:SOUR:FUNC CURR
:SENS:VOLT:PROT 20
:SENS:VOLT:RANG 20
:OUTP:STAT ON
Select source current mode.
Voltage limit when current source is active.
Make sure 1000V range is not active.
Turn source on.
Table 2-11
Current range calibration commands
Step Command/procedure*
Description
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
:SOUR:CURR:RANGE <Range>
:SOUR:CURR -<Source_value>
Take DMM reading.
:CAL:PROT:SOUR <DMM_Reading>
Check 2410 for errors.
:CAL:PROT:SENS <DMM_Reading>
Check 2410 for errors.
:SOUR:CURR 0.0
Take DMM reading.
:CAL:PROT:SOUR <DMM_Reading>
Check 2410 for errors.
:CAL:PROT:SENS <DMM_Reading>
Check 2410 for errors.
:SOUR:CURR +<Source_value>
Take DMM reading.
:CAL:PROT:SOUR <DMM_Reading>
Check 2410 for errors.
:CAL:PROT:SENS <DMM_Reading>
Check 2410 for errors.
:SOUR:CURR 0.0
Take DMM reading.
:CAL:PROT:SOUR <DMM_Reading>
Select source range.
Establish negative full-range polarity.
Read actual output value.
Calibrate sense function negative full scale.
Calibrate source function negative full scale.
Set output to 0A.
Read actual output value.
Calibrate sense function negative zero.
Calibration source function negative zero.
Establish positive full-range polarity.
Read actual output value.
Calibrate sense function positive full scale.
Calibrate source function positive full scale.
Set output to 0A.
Read actual output value.
Calibrate source positive zero.
2-22
Calibration
2410-902-01.book Page 22 Monday, November 7, 2005 2:49 PM
Summary of Contents for Keithley SourceMeter 2410
Page 11: ...1 Performance Verification 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 29: ...2 Calibration 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 53: ...3 Routine Maintenance 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 56: ...4 Troubleshooting 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 74: ...5 Disassembly 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 84: ...6 Replaceable Parts 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 100: ...A Specifications 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 102: ...B Command Reference 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 116: ...C Calibration Programs 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...