Accuracy calculations
The information below discusses how to calculate accuracy for both measurement and source
functions.
Measurement accuracy
Measurement accuracy is calculated as follows:
Accuracy = ±(% of r offset)
As an example of how to calculate the actual reading limits, assume that you are measuring
10V on the 20V range. You can compute the reading limit range from one-year measure voltage
accuracy specifications as follows:
Accuracy = ±(% of r offset)
=
±[(0.015%
×
10V) + 1mV]
=
±(1.5mV + 1mV)
=
±2.5mV
Thus, the actual reading range is 10V± 2.5mV or from 9.9975 to 10.0025V.
Current measurement calculations are performed in exactly the same manner using the perti-
nent specifications, ranges, and input signal values.
Source accuracy
Source accuracy is calculated similarly, except of course that source specifications are used.
As an example of how to calculate the actual source output limits, assume that you are sourcing
0.7mA on the 1mA source range. You can compute the reading limit range from source current
one-year accuracy specifications as follows:
Accuracy = ±(0.034% of 200nA offset)
=
±[(0.034%
×
0.7mA) + 200nA)]
=
±(238nA + 200nA)
=
±438nA
In this case, the actual current output range is 0.7mA ±438nA or from 0.69956mA to
0.70044mA.
A-6
Specifications
2410-902-01.book Page 2 Monday, November 7, 2005 2:49 PM
Summary of Contents for Keithley SourceMeter 2410
Page 11: ...1 Performance Verification 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 29: ...2 Calibration 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 53: ...3 Routine Maintenance 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 56: ...4 Troubleshooting 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 74: ...5 Disassembly 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 84: ...6 Replaceable Parts 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 100: ...A Specifications 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 102: ...B Command Reference 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Page 116: ...C Calibration Programs 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...