Removal and Replacement Procedures
Remove Module From
Platform
1.
Remove the four 1/4” standoff posts.
When reinstalling these standoffs, torque them to 8 in/lb.
2.
Remove the cosmetic cover.
3.
Remove the four 1/4” standoff posts that secure the module to the mainframe
chassis.
When reinstalling these standoffs, torque them to 12 in/lb.
4.
Remove the module from the platform.
CAUTION.
Do not open the module. There are no user-serviceable parts. If
the module is opened, you must return the entire instrument to Tektronix for
calibration.
4–4
H600 & SA2600 Service Manual
Summary of Contents for H600 RFHawk
Page 1: ...xx H600 RFHawk SA2600 Spectrum Analyzer ZZZ Service Manual P071264101 071 2641 01...
Page 2: ......
Page 6: ......
Page 14: ...Service Safety Summary viii H600 SA2600 Service Manual...
Page 17: ...Operating Information...
Page 18: ......
Page 41: ...Theory of Operation...
Page 42: ......
Page 52: ...Theory of Operation 2 10 H600 SA2600 Service Manual...
Page 53: ...Adjustment Procedures...
Page 54: ......
Page 56: ...Adjustment Procedures 3 2 H600 SA2600 Service Manual...
Page 57: ...Maintenance...
Page 58: ......
Page 69: ...Removal and Replacement Procedures H600 SA2600 Service Manual 4 11...
Page 89: ...Replaceable Parts...
Page 90: ......
Page 92: ...Replaceable Parts Figure 5 1 Module and cover replaceable parts 5 2 H600 SA2600 Service Manual...
Page 94: ...Replaceable Parts Figure 5 2 Mainframe replaceable parts 5 4 H600 SA2600 Service Manual...
Page 96: ...Replaceable Parts Figure 5 3 H600 and SA2600 accessories 5 6 H600 SA2600 Service Manual...
Page 98: ...Replaceable Parts 5 8 H600 SA2600 Service Manual...
Page 99: ...Specifications...
Page 100: ......