Performance Veri
fi
cation
5.
Set the Signal Generator as follows:
a.
Reset the Signal Generator.
b.
Set the frequency to
50 MHz
.
c.
Set the output level to
+5 dBm
.
6.
Push the
ZERO
button on the power meter.
7.
Connect the power meter head to the barrel adapter.
8.
Perform a power measurement using the power meter:
a.
Set the power meter correction factor for
50 MHz
.
b.
Read the power.
9.
Adjust the output level of the Signal Generator until the power meter reads
-5 dBm
.
10.
Disconnect the power meter head and the barrel adapter from the cable/10 dB
attenuator assembly.
11.
Connect the RF cable with the 10 dB attenuator to the
RF INPUT
of the
H600 or SA2600 under test.
H600 & SA2600 Service Manual
1–15
Summary of Contents for H600 RFHawk
Page 1: ...xx H600 RFHawk SA2600 Spectrum Analyzer ZZZ Service Manual P071264101 071 2641 01...
Page 2: ......
Page 6: ......
Page 14: ...Service Safety Summary viii H600 SA2600 Service Manual...
Page 17: ...Operating Information...
Page 18: ......
Page 41: ...Theory of Operation...
Page 42: ......
Page 52: ...Theory of Operation 2 10 H600 SA2600 Service Manual...
Page 53: ...Adjustment Procedures...
Page 54: ......
Page 56: ...Adjustment Procedures 3 2 H600 SA2600 Service Manual...
Page 57: ...Maintenance...
Page 58: ......
Page 69: ...Removal and Replacement Procedures H600 SA2600 Service Manual 4 11...
Page 89: ...Replaceable Parts...
Page 90: ......
Page 92: ...Replaceable Parts Figure 5 1 Module and cover replaceable parts 5 2 H600 SA2600 Service Manual...
Page 94: ...Replaceable Parts Figure 5 2 Mainframe replaceable parts 5 4 H600 SA2600 Service Manual...
Page 96: ...Replaceable Parts Figure 5 3 H600 and SA2600 accessories 5 6 H600 SA2600 Service Manual...
Page 98: ...Replaceable Parts 5 8 H600 SA2600 Service Manual...
Page 99: ...Specifications...
Page 100: ......