Performance Veri
fi
cation
Instrument Warmup
Before performing these procedures, the following equipment must have been
operating continuously for at least twenty (20) minutes in an environment that
meets the operating range speci
fi
cations for temperature and humidity:
The H600 or SA2600 under test
All test equipment
Equipment Setup
Setup the test equipment as follows:
1.
Reset all the test instruments to their factory default settings before starting a
test.
2.
Connect the RF cable to the output of the Signal Generator; all test instructions
make this assumption.
3.
Connect the AC adapter to the H600 or SA2600 instrument. Do not attempt to
do the procedure steps while the instrument is powered from the batteries.
H600 & SA2600 Service Manual
1–5
Summary of Contents for H600 RFHawk
Page 1: ...xx H600 RFHawk SA2600 Spectrum Analyzer ZZZ Service Manual P071264101 071 2641 01...
Page 2: ......
Page 6: ......
Page 14: ...Service Safety Summary viii H600 SA2600 Service Manual...
Page 17: ...Operating Information...
Page 18: ......
Page 41: ...Theory of Operation...
Page 42: ......
Page 52: ...Theory of Operation 2 10 H600 SA2600 Service Manual...
Page 53: ...Adjustment Procedures...
Page 54: ......
Page 56: ...Adjustment Procedures 3 2 H600 SA2600 Service Manual...
Page 57: ...Maintenance...
Page 58: ......
Page 69: ...Removal and Replacement Procedures H600 SA2600 Service Manual 4 11...
Page 89: ...Replaceable Parts...
Page 90: ......
Page 92: ...Replaceable Parts Figure 5 1 Module and cover replaceable parts 5 2 H600 SA2600 Service Manual...
Page 94: ...Replaceable Parts Figure 5 2 Mainframe replaceable parts 5 4 H600 SA2600 Service Manual...
Page 96: ...Replaceable Parts Figure 5 3 H600 and SA2600 accessories 5 6 H600 SA2600 Service Manual...
Page 98: ...Replaceable Parts 5 8 H600 SA2600 Service Manual...
Page 99: ...Specifications...
Page 100: ......