Appendix: Warranted Speci
fi
cations
Appendix: Warranted Speci
fi
cations
The limits stated below are for ambient temperature of
≤
30 °C and the bandwidth selection set to FULL.
Reduce the upper bandwidth frequency by 1% for each °C above 30 °C.
Instrument
5 mV/div to 1 V/div
2 mV/div to 4.98 mV/div
1 mV/div to
1.99 mV/div
DPO/MSO4104
DC to 1 GHz
DC to 350 MHz
DC to 200 MHz
DPO/MSO4054
DC to 500 MHz
DC to 350 MHz
DC to 200 MHz
Instrument
2 mV/div to 1 V/div
1 mV/div to 1.99 V/div
DPO/MSO4034
DC to 350 MHz
DC to 200 MHz
Analog bandwidth,
50
Ω
DPO/MSO4032
DC to 350 MHz
DC to 200 MHz
Input impedance, DC
coupled
1 M
Ω
±1% in parallel with 13 pF ±2 pF
50
Ω
±1%
DPO/MSO4101: VSWR
≤
1.5:1 from DC to 1 GHz, typical
DPO/MSO4054: VSWR
≤
1.5:1 from DC to 500 MHz, typical
DPO/MSO4034, DPO/MSO4032: VSWR
≤
1.5:1 from DC to 350 MHz, typical