Analyze Waveform Data
3.
Turn multipurpose knob
a
to select the
speci
fi
c measurement. If needed, then turn
multipurpose knob
b
to select the channel to
measure on.
4.
To remove a measurement, push
Remove
Measurement
, turn multipurpose knob
a
to
select the speci
fi
c measurement, and push
OK Remove Measurement
on the side-bezel
menu.
Quick Tips
To remove all measurements, select
Remove All Measurements
.
A
symbol appears instead of the expected numerical measurement if a vertical clipping condition exists. Part of the
waveform is above or below the display. To obtain a proper numerical measurement, turn the vertical scale and position
knobs to make all of the waveform appear in the display.
MSO4000 and DPO4000 Series Oscilloscopes User Manual
191
Summary of Contents for DPO4032
Page 2: ......
Page 6: ......
Page 12: ...Table of Contents vi MSO4000 and DPO4000 Series Oscilloscopes User Manual...
Page 44: ...Installation 18 MSO4000 and DPO4000 Series Oscilloscopes User Manual...
Page 294: ...Using Application Modules 268 MSO4000 and DPO4000 Series Oscilloscopes User Manual...
Page 344: ...Application Examples 318 MSO4000 and DPO4000 Series Oscilloscopes User Manual...
Page 348: ...Appendix Warranted Specifications 322 MSO4000 and DPO4000 Series Oscilloscopes User Manual...