Get Acquainted with the Instrument
Front-Panel Connectors
1.
Logic Probe Connector (on
MSO4000 models only)
2.
Channel
1
,
2
, (
3
,
4
). Channel inputs with
TekVPI Versatile Probe Interface.
3. Aux In
. Trigger level range is adjustable
from +8 V to –8 V. The maximum input
voltage is 400V peak, 250V RMS. Input
resistance is 1 M
Ω
± 1% in parallel with
13 pF ±2 pF.
4. PROBE COMP
. Square wave signal
source to compensate probes. Output
voltage: 0 – 2.5V, amplitude ± 1% behind
1 k
Ω
±2%. Frequency: 1 kHz.
5.
Ground.
6.
Application Module Slots.
MSO4000 and DPO4000 Series Oscilloscopes User Manual
77