Application Examples
4.
Push
More
repeatedly until you select
Reference Levels
from the popup menu.
Reference
Levels
5.
Push
Set Levels in
to select
units
.
Set Levels
in
% |
Units
6.
Push
High Ref
and turn multipurpose knob
a
to enter 2.00 V. If needed, push
Fine
to change
the sensitivity of the multipurpose knob.
High Ref
(a) 2.00 V
Mid Ref
7.
Push
Low Ref
and turn multipurpose knob
a
to
enter 800 mV. If needed, push
Fine
to change
the sensitivity of the multipurpose knob.
Low Ref
(a) 800 mV
Rise time is typically measured between the 10% and 90% amplitude levels of a signal. These are the default reference levels
that the oscilloscope uses for rise time measurements. However, in this example, you need to measure the time that the signal
takes to pass between the 0.8 V and 2.0 V levels.
You can customize the rise time measurement to measure the signal transition time between any two reference levels. You
can set each of these reference levels to a speci
fi
c percent of the signal amplitude or to a speci
fi
c level in vertical units (such
as volts or amperes).
278
MSO4000 and DPO4000 Series Oscilloscopes User Manual
Summary of Contents for DPO4032
Page 2: ......
Page 6: ......
Page 12: ...Table of Contents vi MSO4000 and DPO4000 Series Oscilloscopes User Manual...
Page 44: ...Installation 18 MSO4000 and DPO4000 Series Oscilloscopes User Manual...
Page 294: ...Using Application Modules 268 MSO4000 and DPO4000 Series Oscilloscopes User Manual...
Page 344: ...Application Examples 318 MSO4000 and DPO4000 Series Oscilloscopes User Manual...
Page 348: ...Appendix Warranted Specifications 322 MSO4000 and DPO4000 Series Oscilloscopes User Manual...