Performance veri
fi
cation
Table 41: Basic Mode Harmonic Distortion Test Record (cont.)
Harmonic Distortion
(amplitude 1 V
p-p
)
Fundamental
= reference
2nd
3rd
4th
5th
Limit
Sine 100 MHz
100 MHz
200 MHz
300 MHz
400 MHz
500 MHz
CH2+ Harmonic
Distortion (dBc)
reading -
reference
(dBc)
0 dBc
Nth - reference <
-40 dBc
CH2– Harmonic
Distortion (dBc)
reading -
reference
(dBc)
0 dBc
Nth - reference <
-40 dBc
Sine 300 MHz
300 MHz
600 MHz
900 MHz
1200 MHz
1500 MHz
CH2+ Harmonic
Distortion (dBc)
reading -
reference
(dBc)
0 dBc
Nth - reference <
-28 dBc
CH2– Harmonic
Distortion (dBc)
reading -
reference
(dBc)
0 dBc
Nth - reference <
-28 dBc
Sine 600 MHz
600 MHz
1200 MHz 1800 MHz
2400 MHz
3 GHz
CH2+ Harmonic
Distortion (dBc)
reading -
reference
(dBc)
0 dBc
Nth - reference <
-28 dBc
CH2– Harmonic
Distortion (dBc)
reading -
reference
(dBc)
0 dBc
Nth - reference <
-28 dBc
AWG4162 Speci
fi
cations and Performance Veri
fi
cation
53