Performance veri
fi
cation
The following types of performance veri
fi
cation procedures can be performed on
this product. You may not need to perform all of these procedures, depending on
what you want to accomplish.
Self tests: Perform to quickly con
fi
rm that the instrument is operating
properly. These tests take little time and check the overall functionality of the
internal hardware of the instrument.
Performance tests: Perform to directly check warranted speci
fi
cations. These
tests require additional equipment and time to execute than the self tests.
NOTE.
The performance veri
fi
cation procedures are not calibration procedures.
The performance veri
fi
cation procedure only veri
fi
es that your instrument meets
key speci
fi
cations. For your instrument to be calibrated, it must be returned to a
Tektronix service facility.
Prerequisites
All of the tests in this section con
fi
rm performance and functionality when the
following requirements are met:
The instrument is operating in an environment that meets the temperature,
altitude, and humidity characteristics listed in the speci
fi
cations.
The instrument is completely assembled and covers installed per factory
speci
fi
cation.
The instrument has been operating for a warm-up period of at least for
30 minutes.
The instrument must be operating at an ambient temperature b5
°C and +50 °C.
Performance conditions
In addition to the above requirements, all of the performance tests in this section
comprise an extensive, valid con
fi
rmation of performance and functionality when
the following additional requirement is met:
AWG4162 Speci
fi
cations and Performance Veri
fi
cation
25