Performance veri
fi
cation
Basic mode AC Flatness Test Record (cont.)
Frequency 400.00 MHz
Reference - 0.5 dB
dB
Ref 0.50 dB
Frequency 500.00 MHz
Reference - 0.5 dB
dB
Ref 0.50 dB
Frequency 600.00 MHz
Reference - 0.5 dB
dB
Ref 0.50 dB
CH2– AC Flatness
Minimum
Test result
Maximum
Frequency 1.00 kHz
(Ampl: +1 Vpp)
———-
dB ( = Reference)
———-
Frequency 500 kHz
Reference - 0.5 dB
dB
Ref 0.50 dB
Frequency 1.00 MHz
Reference - 0.5 dB
dB
Ref 0.50 dB
Frequency 5.00 MHz
Reference - 0.5 dB
dB
Ref 0.50 dB
Frequency 10.00 MHz
Reference - 0.5 dB
dB
Ref 0.50 dB
Frequency 50.00 MHz
Reference - 0.5 dB
dB
Ref 0.50 dB
Frequency 100.00 MHz
Reference - 0.5 dB
dB
Ref 0.50 dB
Frequency 150.00 MHz
Reference - 0.5 dB
dB
Ref 0.50 dB
Frequency 200.00 MHz
Reference - 0.5 dB
dB
Ref 0.50 dB
Frequency 300.00 MHz
Reference - 0.5 dB
dB
Ref 0.50 dB
Frequency 400.00 MHz
Reference - 0.5 dB
dB
Ref 0.50 dB
Frequency 500.00 MHz
Reference - 0.5 dB
dB
Ref 0.50 dB
Frequency 600.00 MHz
Reference - 0.5 dB
dB
Ref 0.50 dB
Table 41: Basic Mode Harmonic Distortion Test Record
Harmonic Distortion
(amplitude 1 V
p-p
)
Fundamental
= reference
2nd
3rd
4th
5th
Limit
AWG4162
Spectrum Analyzer reading
Sine 5 MHz
5 MHz
10 MHz
15 MHz
20 MHz
25 MHz
CH1+ Harmonic
Distortion (dBc)
reading -
reference
(dBc)
0 dBc
Nth - reference <
-60 dBc
CH1– Harmonic
Distortion (dBc)
reading -
reference
(dBc)
0 dBc
Nth - reference <
-60 dBc
Sine 100 MHz
100 MHz
200 MHz
300 MHz
400 MHz
500 MHz
CH1+ Harmonic
Distortion (dBc)
AWG4162 Speci
fi
cations and Performance Veri
fi
cation