Performance veri
fi
cation
3.
Connect the generator to the DMM as shown in the following
fi
gure. Be sure
to connect the 50
Ω
terminator to the generator’s side output connector.
Figure 6: DC offset/Vocm tests, Basic mode
4.
Set up the generator using the following steps:
Select menu
Setting
Operation
Function
DC
Home > More > DC
Offset
1 V
Home > Offset
Channel 1 Output
On
On (front)
5.
Verify that each offset measurement is within the range speci
fi
ed in the test
record.
6.
Record the results in the test record.
7.
Repeat steps 4 through 6 for the CH1–, CH2+, and CH2– outputs.
8.
Test the V
ocm
as follows:
a.
Set DC offset to 0 V (Home > More > DC > Offset).
b.
Set up the generator using the following steps:
Select menu
Setting
Operation
Function
Vocm
System > Output CH1
Vocm
1 V
System > Output CH1 > Vocm
Channel 1 Output
On
On (front)
9.
Verify that each Vocm measurement is within the range speci
fi
ed in the test
record.
10.
Record the results in the test record.
11.
Repeat steps 8 through 10 for the CH1–, CH2+, and CH2– outputs.
AWG4162 Speci
fi
cations and Performance Veri
fi
cation
37