Characteristic
TRIGGER SENSITIVITY
Intl Ac Norm
Intl Ac Auto
Ext De
TRIGGER JITTER
INPUT IMPEDANCE
MAX SAFE INPUT VOL TAGE
1
MAX USABLE INPUT VOL TAGE
1
1 Supplemental information only.
40
TABLE 3 (cont.)
Oscilloscope Electrical Specifications
Performance Requirement
TRIGGERING
0.5 div, 7 Hz to 1 MHz
Same as Intl Ac Norm except sweep free runs when input
signal is not sufficient to cause triggering
1 V, de to 1 MHz
0.5 µs or less at 1 MHz
1 MS1 within 10% shunted by approximately 20 pF
200 V (de + peak ac)
±8 V peak at 1 MHz or less
213