TABLE 2 (continued)
Digital Multimeter Electrical Specifications
Characteristic
Performance Requirement
RESISTANCE
Range
1 kS1 to 10 MD
Resolution
1 S1 on 1 kS1 range
Accuracy (20
Q
C to 30
Q
C)
Refer to opening paragraphs under Operators Specifications
(Batteries only)
for power I ine operation
1 kS1
Within 0.5% of reading± 3 counts. Temp Coef is (within
0.03% of 0.04% of full scale) per
Q
C
10 kS1 to 1 MD
Within 0.5% of reading± 1 count. Temp Coef is (within
0.02% of 0.02% of full scale) per °C
10 MD
Within 1% of reading± 1 count. Temp Coef is (within 0.05%
of 0.02% of full scale) per QC
Accidental Voltage
_
I
nput Protection
125 V maximum on all ranges
Settling Time
2 seconds for reading within 2 counts
Reference Current 1
Approximately 100 µA, 100 µA, 10 µA, 1 µA, and
100
nA
for ranges 1 kS1through 10 MD respectively
1
Supplemental information only.
REV. A, MAR. 1975