TABLE 2 (continued)
Digital Multimeter Electrical Specifications
Characteristic
Performance Requirement
DC CURRENT
@
Accuracy (+20
°
C to +30
°
C)
(Batteries only-refer to opening paragraphs un
der Operators Specifications for power I ine
operation)
0.1 mA Range
1 mA to 1000 mA
Settling Time
Temp Coef is (within 0.02% of r0.04% of full scale)
per
°
C
Within 0.5% of reading± 3 counts
Within 0.25% of readinq ± 3 counts
1.5 seconds for reading within 0.1 %