TABLE 2 (continued)
Digital Multimeter Electrical Specifications
Characteristic
Performance Requirement
DC VOLTS
Accuracy (+20
°
C to +30
°
C) ( Batteries Only)
Refer to the opening paragraphs under Operators Specifica-
tions for power I ine operation
0.1 V
Within 0.1% of reading± 3 counts. Temp Coef is (within
0.015% of 0.04% of full scale) per °C.
1 V
Within 0.1% of reading± 1 count. Temp Coef is (within
0.01 % of 0.01 % of full scale) per °C.
10 V and 100 V
Within 0.15% of reading ±1 count. Temp Coef is (within
0.015% of 0.01% of full scale) per
°
C
1000 V
Within 0.2% of reading± 1 count. Temp Coef is (within
0.02% of 0.01% of full scale) per
°
C
Settling Time
1.5 seconds for reading with in 0.1 %
Normal Mode Rejection Ratio
-60 dB at 60 Hz, -3 dB at 1 Hz
30