TABLE 2 (continued
Digital Multimeter Electrical Specifications
Characteristic
Performance Requirement
RMS VOLTS
Accuracy (20
°
C to +30
°
C) (Batteries only - refer to
Temp Coef is (within 0.05% of r0.1 % of full scale)
opening paragraphs under Operators Specifications for
per
°
C
power line operation). Crest factor< 2, reading at
least 10% of full scale.
Within% of reading shown± 5 counts at frequency shown:
De
40 Hz
4 kHz
to
to
4 kHz
40 kHz
0.1 V
2.5%
1.5%
3.5%
1 V
2%
1%
1%
10 V
2%
1%
1%
100 V
2%
1%
1%
1000 V
2%
1%
2%
Settling Time
2 seconds for reading within 1%
Max Crest Factor
1
Approximately 5
1 Supplemental information only.
@