How to adjust the distance value to make the device measure level correctly
•
Go to the signal screen in operator mode. For more data on how to get to the signal screen,
refer to
Operator mode
on page 65.
•
Push [>>>>] one time to move the cursor right to select the level signal
1
.
•
Make a note of the distance value on the signal screen.
•
Push [
^
^
^
^
] for three seconds.
i
The program mode screen is displayed.
•
Move the cursor to Supervisor
Supervisor
Supervisor
Supervisor and push [
^
^
^
^
]
2
.
i
Type in the supervisor password, if this is necessary. For more data, refer to
How to get
access to the supervisor mode
on page 68.
•
Type in the distance value
3
.
i
The selected signal is the level signal.
•
Push [
^
^
^
^
] to accept your new value or push [>>>>] and [
] (Esc) at the same time to reject it.
i
The display goes back to the signal screen.
How to adjust the interface distance value to make the device measure interface
correctly (devices for interface measurement only).
•
Go to the signal screen in operator mode. For more data, refer to
Operator mode
on page 65.
•
Push [>>>>] 2 times to move the cursor right to select the interface signal.
•
Make a note of the distance value on the signal screen.
•
Push [
^
^
^
^
] for three seconds.
i
The program mode screen is displayed.
Figure 6-7: Areas on the signal screen
1
Reference signal area (the value cannot be changed)
2
Level signal area (use the distance input function to change the level value)
3
Interface signal area (use the distance input function to change the interface value)
4
Probe end area (in TBF mode only
–
this value cannot be changed)
Figure 6-8: How to make the device follow the correct level signal
88