Installation in tanks containing one liquid and foam
•
Drill a pressure equalization hole in the stilling well above the maximum level.
•
Deburr the hole.
•
If the probe has a counterweight, make sure that there is enough space between the
counterweight and the wall of the stilling well.
Installation in tanks containing more than one liquid
•
Drill a pressure equalization hole in the stilling well above the maximum level of the top liquid.
•
Drill more holes along the length of the stilling well. Distance between holes
≥
25 mm / 1
¨
(depends on the minimum layer to be measured)
i
These holes help the liquids to move freely.
•
Deburr the holes.
•
If the probe has a counterweight, make sure that there is enough space between the
counterweight and the wall of the stilling well.
Floating roofs
If the device is for a tank with a floating roof, install it in a stilling well.
Figure 3-18: Installation recommendations for stilling wells
1
Stilling well in tanks containing one liquid
2
Stilling well in tanks containing more than one liquid
3
Air circulation hole
4
Maximum level of the liquid
5
Liquid circulation hole
6
Distance between holes
≥
25 mm/1
¨
(depends on the minimum layer to be measured)
2
6
www.tek-trol.com
Technology Solutions
Instruction Manual
Tek-Flex
4100A