Regardless of the input connection type, i.e. connector or flying lead, the following electrical input signal types may be
attached to each of the 'General Purpose inputs (GPis)':
Digital: Volt-free contacts or logic pulses
Count (pulses counted over, and recorded at, pre-set intervals)
State / Tamper (status and time of state change recorded)
Event (time of pulse recorded)
Frequency (switch closure or logic pulses, maximum frequency = 16kHz, programmable sampling period of 1 to 250
seconds, independent of recording rate)
Analogue: 0 - 2.5Vdc, 0.01V resolution
Notes
For 'digital' inputs, '0' < 0.5Vdc and '1' > 2Vdc.
For 'analogue' inputs, 'Low' < 0.5Vdc and 'High' > 2Vdc.
All options dependent on configuration file.
Refer also to the "Standard connections: Channel mapping & channel shadowing section of this document.
o
o
1.
2.
3.
•
Optional open-face sealed, polarised / keyed 3-way 'Mil.
Spec. compatible' bayonnet fitting female connector
Note - as indicated above, a cable from the Technolog
CBL001-x range may be used to attach to this connector.
Optional 'flying lead' input cable
Flying lead inputs
Cable core
Function
Black
GPi3
Green
GPi2
Blue
GND (0V)
Connector inputs
Pin
Function
CBL001-x
A
GPi3
Black
B
GPi2
White
C
GND (0V)
Red
Standard connections - Port 3:
Two signal input (analogue and / or digital) connection