![Technogym SYNCHRO Excite + Service And Maintenance Manual Download Page 188](http://html1.mh-extra.com/html/technogym/synchro-excite/synchro-excite_service-and-maintenance-manual_1070236188.webp)
SYNCHRO : Service & maintenance manual - rev. 7.1
Pagina 11.4
11.4.
LIST OF CRITICAL SPARE PARTS
Technogym medical devices are identified by “Serial N” plates that allow them to be tracked
throughout the useful life-span of the device. Just as for products, there is also a list of so-called
CRITICAL
components which, through the SN, must be guaranteed as traceable.
The list of these critical components is shown in the table below:
Code
Component
R0006668XX
Brake Board
0WR00503XX
Solenoid assembly
R0003767XX
Display Board 700
R0003778XX
Display Board 500
0WQ00060XX
C-Safe Board
0WR00633XX
Cardio Receiver HR/HS
N0000034XX
Cardio Receiver HR
W0005566XX
Schuster socket
0WR00598XX
N. 2 ferrites RKCF-10-A5 RICHCO
-
Display program – SW version
-
Brake board program – SW version
Note:
XX = Baan Revision
YY = Languages
If a critical component listed in this table is replaced during a repair, maintenance or refurbishment,
this fact must be
noted
in the
SERVICE REPORT,
by recording the
tracking information
for the
new component.
Summary of Contents for SYNCHRO Excite +
Page 1: ...S Sy yn nc ch hr ro o E X C I T E REV 7 1 ...
Page 2: ......
Page 4: ......
Page 10: ...SYNCHRO Excite Service maintenance manual rev 7 1 Page vi Page intentionally left blank ...
Page 50: ...SYNCHRO Excite Service maintenance manual rev 7 1 Page 3 18 Page intentionally left blank ...
Page 98: ...SYNCHRO Excite Service maintenance manual rev 7 1 Page 6 40 Page intentionally left blank ...
Page 182: ...SYNCHRO Excite Service maintenance manual rev 7 1 Page 9 26 Page intentionally left blank ...
Page 184: ...SYNCHRO Excite Service maintenance manual rev 7 1 Pagina 10 2 Page intentionally left blank ...
Page 195: ......