Proline Promass 300 EtherNet/IP
Commissioning
19
4.3.3
Via Ethernet-based fieldbus
If the IP address of the measuring device is assigned via hardware or software addressing,
the network connection can be established directly via the Ethernet network.
The device is configured at the factory to work with the Dynamic Host Configuration
Protocol (DHCP), i.e. the IP address of the measuring device is automatically assigned
by the automation system (DHCP server). To establish a network connection via the
service interface (CDI-RJ45): the "Default Ethernet network settings" DIP switch must
be set to
ON
. The measuring device then has the fixed IP address: 192.168.1.212.
This address can now be used to establish the network connection → 16.
4
2
1
3
5
5
5
A0016961
2
Ethernet network
1
Automation system, e.g. "RSLogix" (Rockwell Automation)
2
Workstation for measuring device operation: with Add-on Profile Level 3 for "RSLogix 5000" (Rockwell
Automation) or with Electronic Data Sheet (EDS)
3
Computer with Web browser (e.g. Internet Explorer) for accessing the integrated device Web server or with
operating tool (e.g. FieldCare, DeviceCare) with COM DTM "CDI Communication TCP/IP"
4
Ethernet switch
4.4
Establishing a connection to the Web server
4.4.1
Prerequisites
The IP settings in the measuring device and computer must match before a connection can
be established successfully. In particular this concerns the IP addressing and Web browser
settings.
The following conditions must be met to connect:
• The Web server of the measuring device is enabled → 15.
• The IP address of the measuring device is known → 15.
• The computer used meets the requirements for hardware and software → 9.
• The measuring device and computer are connected to one another → 16
• The measuring device is switched on.